0
PRODUCTS PRODUCTS INDEX
0
 DC MINI-MOTORS
COMPONENTS
MAP-HD-140
Micro Air Pump Motor
| HTML | PDF | DIMENSIONS |
0
photo 0
[APPLICATIONS]
Blood Pressure monitor for Wrist type

[SPECIFICATIONS]
Items
Specifications
Conditions
  Rated Voltage
3.0V
 
  Voltage Range
2.0~3.2V
 
  Pressing Time
10.0sec or less
Time to add pressure from 0 to 300mmHg to a tank of 100cc by rated voltage (DC 3V).
  Current Consumption  
450mA or less
Maximum current during adding pressure from 0 to 300mmHg to a tank of 100cc by rated voltage (DC 3V).
  Leakage of Air
3.0mmHg/min or less
Decrease of pressure at 30 second after stop pumping to a tank of 100cc up to 40 ± 10mmHg and 300 ± 10mmHg.
  Maximum Pressure
350mmHg or more
 
  Noise
55.0dB or less
Noise at 30cm from the pump.
0
[NOTE]
-Any products mentioned in this catalog are subject to any modification in their appearance and others for improvements without prior notification.
-The details listed here are not a guarantee of the individual products at the time of ordering. When using the products, you will be asked to check their specifications.
0

PAGE TOP
BACK